DICOM PS3.4 2017d - Service Class Specifications
V.6.2.2.2 Substance Approval Query Attributes

Table V.6-2 defines the Attributes of the Substance Approval Query Information Model.

Table V.6-2. Attributes for the Substance Approval Query Information Model

Description / Module

Tag

Matching Key Type

Return Key Type

Remark/Matching Type

Patient

Patient's Name

(0010,0010)

O

2

Patient ID

(0010,0020)

R

1

Shall be retrieved with Single Value Matching only.

One or both of Patient ID (0010,0020) and Admission ID (0038,0010) shall be present as a Matching Key in the Query

Issuer of Patient ID

(0010,0021)

O

2

Issuer of Patient ID Qualifiers Sequence

(0010,0024)

O

3

>All Attributes of the Issuer of Patient ID Qualifiers Sequence

O

3

Patient's Birth Date

(0010,0030)

-

2

Patient's Sex

(0010,0040)

-

2

Visit

Admission ID

(0038,0010)

R

2

Shall be retrieved with Single Value Matching only.

One or both of Patient ID (0010,0020) and Admission ID (0038,0010) shall be present as a Matching Key in the Query

Issuer of Admission ID Sequence

(0038,0014)

O

3

>Local Namespace Entity ID

(0040,0031)

O

1C

Required if Universal Entity ID (0040,0032) is not present; may be present otherwise

>Universal Entity ID

(0040,0032)

O

1C

Required if Local Namespace Entity ID (0040,0031) is not present; may be present otherwise.

>Universal Entity ID Type

(0040,0033)

O

1C

Required if Universal Entity ID (0040,0032) is present.

Product

Product Package Identifier

(0044,0001)

R

1

Shall be retrieved with Single Value Matching only. Shall be present as a Matching Key in the Query.

Substance Administration

Administration Route Code Sequence

(0054,0302)

R

1

Shall be present as a Matching Key in the Query.

>Code Value

(0008,0100)

R

1

>Coding Scheme Designator

(0008,0102)

R

1

>Code Meaning

(0008,0104)

-

1

Approval

Substance Administration Approval

(0044,0002)

-

1

Approval Status Further Description

(0044,0003)

-

2

Approval Status DateTime

(0044,0004)

-

1


One or both of Patient ID (0010,0020) and Admission ID (0038,0010) shall be present as a Matching Key in the Query.

Product Package Identifier (0044,0001) shall be present as a Matching Key in the Query. The Product Package Identifier might not be globally unique and might conflict with other identifiers used within the scope of the institution.

Note

The package identifiers are typically unique within the scope of the substance administration management systems. This is a warning that they are not UIDs.

Administration Route Code Sequence (0054,0302) shall be present as a Matching Key in the Query, and a single Item shall be present in that Sequence with Code Value (0008,0100) and Coding Scheme Designator (0008,0102) as Matching Keys.

DICOM PS3.4 2017d - Service Class Specifications