DICOM PS3.4 2017d - Service Class Specifications

P.3.2 Operation

The DICOM AEs that claim conformance to this SOP Class as an SCU shall invoke the N-ACTION request. The DICOM AEs that claim conformance to this SOP Class as an SCP shall support the N-ACTION request.

P.3.2.1 Substance Administration Log Action Information

This operation allows an SCU to submit a Medication Administration Record log item or entry, providing information about a specific real-world act of Substance Administration that is the purview of the SCU. This operation shall be invoked through the DIMSE N-ACTION Service.

The Action Information Attributes are defined by the Substance Administration Log Module specified in PS3.3. The DICOM AEs that claim conformance to this SOP Class as an SCU and/or an SCP shall support the Action Type and Action Information Attributes in the N-ACTION-RQ as specified in Table P.3-2.

Table P.3-2. Substance Administration Logging N-ACTION Information

Action Type Name

Action Type ID

Attribute Name

Tag

Requirement Type SCU/SCP

Record Substance Administration Event

1

Specific Character Set

(0008,0005)

1C/1C

(Required if an extended or replacement character set is used)

Patient ID

(0010,0020)

1C/1C

Either or both Patient ID and Admission ID shall be supplied by the SCU; the SCP shall support the Attribute if supplied

Issuer of Patient ID

(0010,0021)

3/2

Patient's Name

(0010,0010)

2/2

Admission ID

(0038,0010)

1C/1C

Either or both Patient ID and Admission ID shall be supplied by the SCU; the SCP shall support the Attribute if supplied

Issuer of Admission ID

(0038,0011)

3/2

Product Package Identifier

(0044,0001)

1C/1C

Either or both Product Package Identifier and Product Name shall be supplied by the SCU; the SCP shall support the Attribute if supplied

Product Name

(0044,0008)

1C/1C

Either or both Product Package Identifier and Product Name shall be supplied by the SCU; the SCP shall support the Attribute if supplied

Product Description

(0044,0009)

3/3

Substance Administration DateTime

(0044,0010)

1/1

Substance Administration Notes

(0044,0011)

3/2

Substance Administration Device ID

(0044,0012)

3/3

Administration Route Code Sequence

(0054,0302)

2/2

>Code Value

(0008,0100)

1/1

>Coding Scheme Designator

(0008,0102)

1/1

>Code Meaning

(0008,0104)

1/1

Substance Administration Parameter Sequence

(0044,0019)

3/3

> All Attributes of the Substance Administration Parameter Sequence

3/3

Operator Identification Sequence

(0008,1072)

1/1

>Person Identification Code Sequence

(0040,1101)

1/1

>>Code Value

(0008,0100)

1/1

>>Coding Scheme Designator

(0008,0102)

1/1

>>Code Meaning

(0008,0104)

1/1


DICOM PS3.4 2017d - Service Class Specifications