DICOM PS3.16 2017e - Content Mapping Resource

TID 10005 Accumulated Mammography X-Ray Dose

This modality specific Template provides detailed information on breast imaging projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).

Type:

Extensible

Order:

Significant

Root:

No

Table TID 10005. Accumulated Mammography X-Ray Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (111637, DCM, "Accumulated Average Glandular Dose")

1-2

M

UNITS = EV (mGy, UCUM, "mGy")

2

>

HAS CONCEPT MOD

CODE

EV (G-C171, SRT, "Laterality")

1

M

DCID 6022 “Side”


DICOM PS3.16 2017e - Content Mapping Resource